General Introduction
Yokogawa GD402 gas density analyzer converter works with detector GD40 for measuring gas density, specify gravity and molecular weight, it is a rugged microprocessor based converter designed in two versions to meet both general area and explosion-proof application requirements. In addition to the display of several key data items, the converter also provides the choice of three different means for calibration: automatic; semiautomatic and one-touch manual operation. GD402 gas density analyzer converter comes with a simple, and user-friendly interface with low installation cost.
General Specification
Main Characters
|
Description
|
Output Signals
|
4-20mA DC, 2 outputs
|
Power Supply
|
100 to 240 V AC, 24 V DC
|
Power Consumption
|
Approximately 12 W
|
Sample gas conditions
|
All gases except for corrosive gas and acetylene gas
|
Sample gas temperature
|
-10 to 50°C (non-condensing)
|
Sample gas pressure
|
50kPa to 588.4kPa (abs)
|
Sample gas flow
|
0.1 to 1 L/min
|
Construction
|
GD402G: General purpose converter. (Non-Explosion-proof)
|
GD402T: FM Explosion-proof Approval.
|
GD402V: CSA Explosion-proof Approval
|
GD402R: TIIS Explosion-proof Approval
|
3.Measure Specification
Item
|
Density kg/m3
|
Density lb/ft3
|
Specific Gravity
|
Molecular Weight
|
Concentration vol%
|
Range
|
0 - 6 (compensated)
0 - 60 (physical)
|
0 - 0.4 (compensated)
0 - 4 (physical)
|
0 - 5
|
0 - 140
|
0 - 100
|
Minimum Range
|
0.1
|
0.01
|
0.1
|
4
|
Concentration equivalent
to 0.100 kg/m3
|
Response Time 90%
|
approx. 5 sec
|
approx. 5 sec
|
approx. 5 sec
|
approx. 5 sec
|
approx. 5 sec
|
Linearity
|
+/-1%FS
|
+/-1%FS
|
+/-1%FS
|
+/-1%FS
|
+/-1
|
Repeatability
|
+/-0.001
or +/-0.5%FS
|
+/-0.0001
or +/-0.5%FS
|
+/-0.001
or +/-0.5%FS
|
+/-0.02
or +/-0.5%FS
|
+/-0.5% or Concentration
equivalent to
+/-0.001kg/m3
|
Long term stability
|
+/-0.003/month
|
+/-0.002/month
|
+/-0.003/month
|
+/-0.07/month
|
Concentration equivalent
to +/-0.003 kg/m3/month
|
4.Model Selection
Model
|
Suffix Codes
|
Description
|
GD402G
|
|
|
|
|
General purpose model, 6 cable glands included
|
GD402T
|
|
|
|
|
FM certified explosion proof model. Gland threads 1/2 NPT. No cable glands included
|
GD402V
|
|
|
|
|
CSA certified explosion proof model. Gland threads 1/2 NPT. No cable glands included
|
GD402R
|
|
|
|
|
TIIS certified explosion proof model. Gland threads G3/4. No cable glands included
|
Label approval
|
-D
|
|
|
|
24 V DC
|
-A
|
|
|
|
100-240 V AC
|
Label and approval
|
|
-E
|
|
|
English label
|
|
-J
|
|
|
TIIS approval, English label (only GD402R)
|
Instruction Manual
|
|
|
-E
|
|
English
|
Options
|
|
|
|
/PA
|
Panel mounting
|
|
|
|
/U
|
Universal (Pipe and Wall) Mounting
|
|
|
|
/H
|
Hood
|